Troubleshooting - Panasonic MDF-U4186S Series Operating Instructions Manual

Ultra-low temperature freezer
Table of Contents

Advertisement

TROUBLESHOOTING

If the unit malfunctions, check out the following before calling for service.
Malfunction
The chamber is not cooled
at all
The cooling is poor
Alarm test key cannot
actuate the alarm
Note:
If the malfunction is not eliminated after checking the above items, or the malfunction is not shown in the
above table, contact our sales representative or agent.
The voltage is too low (In this case, call an electrician).
The power is not supplied.
The breaker is free.
The ambient temperature is too high.
The door is not shut tightly.
The inner door is not installed correctly.
The set temperature in the controller is not set properly.
The grille is blocked out.
The filter is clogged.
The freezer is in the direct sunlight.
There is any heating source near the freezer.
A rubber cap and insulation are not set correctly.
You put too many unfrozen articles into the chamber.
The alarm is activated only when the power switch is ON.
When only the buzzer or only the alarm is actuated by the alarm test
key, the unactuated part is out of order, and must be replaced.