% 4 – 20mA MEASUREMENTS
1.
Set up and connect as described for DC mA measurements.
2.
Set the rotary function switch to the 4-20mA% position.
3.
The meter will display loop current as a % with 0mA=-25%, 4mA=0%, 20mA=100%, and
24mA=125%.

AUTORANGING/MANUAL RANGE SELECTION

When the meter is first turned on, it automatically goes into AutoRanging. This automatically selects
the best range for the measurements being made and is generally the best mode for most
measurements. For measurement situations requiring that a range be manually selected, perform
the following:
1.
Press the RANGE key. The "AUTO" display indicator will turn off.
2.
Press the RANGE key to step through the available ranges until you select the range you want.
3.
To exit the Manual Ranging mode and return to Autoranging, press EXIT
Note: Manual ranging does not apply for the Temperature functions.
MAX/MIN
1.
Press the MAX/MIN key to activate the MAX/MIN recording mode. The display icon "MAX" will
appear. The meterleft auxiliary displaywill display and hold the maximum reading and will
update only when a new "max" occurs.The display icon "MIN" will appear. The right auxiliary
displaymeter will display and hold the minimum reading and will update only when a new "min"
occurs.
2.
To exit MAX/MIN mode press EXIT

RELATIVE MODE

The relative measurement feature allows you to make measurements relative to a stored reference
value. A reference voltage, current, etc. can be stored and measurements made in comparison to
that value. The displayed value is the difference between the reference value and the measured
value.
1.
Perform the measurement as described in the operating instructions.
2.
Press the REL button to store the reading in the display and the "▲" indicator will appear on
the display.
3.
The Right auxiliary display displays the initial reading (the stored value)
4.
The Left auxiliary display displays the actual currently measured value.
5.
Main display displays the Relative value (currently measured value minus the stored value).