Calibration - Health O Meter 2842KL User Instructions

Large platform digital scale
Hide thumbs Also See for 2842KL:
Table of Contents

Advertisement

CALIBRATION

This scale has been factory calibrated, and does not require calibration prior to use. If
required, the scale can be calibrated. The following calibration assumes the scale is set to
pounds, and 200 lb is employed as the calibration standard.
Note: Only weights certified and traceable to national standards are to be used for
calibration procedures.
Scale Calibration
Press and hold the REWEIGH and ZERO buttons at the same
time, while pressing the ON/OFF button to turn on the scale.
"LOAD" will briefly appear on the display, followed by "200.0
LB" indicates you can calibrate. Release all three buttons.
Ensure there is no weight on the scale (except for the rubber
mat) and press the REWEIGH button. "Clear" will flash on the
display, followed by "--------". At this time the scale is
calculating zero calibration. After zero calibration, the display
will flash "PUT" followed by "200.0 LB".
Place 200 lb on the scale. Then press REWEIGH. "CAL" will
appear on the display, followed by "FAC" followed by 5 digits.
The calibration factor should be in the range between 8000
and 10000 If the number is out of that range, contact
Customer Service at 1-800-815-6615.
If the calibration factor is within range, press the REWEIGH
button. "DONE" will appear on the display, followed by
"START", followed by "0.0". Remove the weight from the
scale.
Calibration procedure is complete.
ACTION
14
DISPLAY SHOWS
"LOAD" followed by
"200.0 LB"
"Clear" followed by
"-------". Then "PUT"
followed by "200.0 LB"
"CAL" followed by "FAC"
followed by "X.XXXX".
"DONE" followed by
"START", followed by
"0.0".

Hide quick links:

Advertisement

Table of Contents
loading