Sanitas SBM 42 Instructions For Use Manual page 38

Table of Contents

Advertisement

Available languages

Available languages

Table 3
For ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacturer's declaration – electromagnetic immunity
The BM 58 is intended for use in the electromagnetic environment specified below. The customer or the user of the BM 58 should assure that
it is used in such an environment.
IMMUNITY test
IEC 60601
test level
Radiated RF
3 V/m 80
IEC 61000-4-3
MHz to 2.5
GHz
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from struc-
tures, objects and people.
Compliance
Electromagnetic environment – guidance
level
Portable and mobile RF communications equipment should be used no closer to any part
of the SBM 42, including cables, than the recommended separation distance calculated
from the equation applicable to the frequency of the transmitter.
Recommended separation distance:
3 V/m
d
P
1
2 .
d
2
3 .
P
Where P is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer and d is the recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site
survey,
should be less than the compliance level in each frequency range
a
Interference may occur in the vicinity of equipment
marked with the following symbol:
80 MHz to 800 MHz
800 MHz to 2,5 GHz
38
.b

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents