1x Instruction Manual Signed by: Date: Major Science is liable for all missing or damaged parts / accessories within 7 days after customer received this instrument package. Please contact Major Science immediately regarding this issue. If no response within such time period from consignee party, that will be consignee...
Section 1 Warning Major Science Mini Pro 300V Power Supply has been tested and found to comply with the limits for the CE regulation. Also, it is RoHS compliant to deliver confident product which meets the environmental directive. These limits are designed to provide reasonable protection against harmful interference when the instrument series is operated in a commercial environment.
Page 5
Avoiding Electrical Shock Follow the guidelines below to ensure safe operation of the unit. The Mini Pro 300V Power Supply has been designed to utilize shielded wires thus minimizing any potential shock hazard to the user. Major Science recommends against the use of unshielded wires.
Page 6
Avoiding Damage to the Instrument 1. Do not attempt to operate the device if damage is suspected. 2. Protect this unit from physical damage, corrosive agents and extreme temperatures (direct sunlight, etc.). 3. For proper ventilation and safety concerns, keep at least 10 cm of space behind the instrument, and at least 5 cm of space on each side.
Page 7
Max. voltage: 300 V Max. current: 400 mA Max. watt: 60 W This product has been tested to the requirements of CAN/CSA-C22.2 No. 61010-1, second edition, including Amendment 1, or a later version of the same standard incorporating the same level of testing requirements THIS LIMITED WARRANTY SHALL NOT APPLY IF THE DEFECT WAS CAUSED THROUGH ANY OF THE FOLLOWING...
Page 8
of dirt or dust, or mains supply problem, thunderstorm activities, or infestation by insects or vermin, or exposure to abnormally corrosive conditions, or operation with extreme heat or humidity.
Section 2 Introduction 2.1 Overview Major Science Mini Pro 300V Power Supplies are recognized as unique power supplies equipped with very powerful specifications to cover the majority of electrophoresis applications on the market. Sufficient and accurate output voltages, two terminal pairs, and its compacted size can deliver accurate and reliable experimental results from one experiment to another.
2.2 Product Description & Feature Combining small size and versatility, the newly redesigned Mini Pro 300V power supply is an ideal choice for any researcher. Capable of providing constant current or constant voltage in 1 mA or 1 V steps, the unit is perfectly suited to run both vertical polyacrylamide and horizontal agarose electrophoresis experiments.
Section 3 Technical Specification Cat. No MINI-300 Output Voltage / Inc. 10 - 300V / 1V Output Current / Inc. 10 - 400mA / 1mA Max. Watt Output Type Constant Voltage or Constant Current Control Microprocessor controller Terminal Pairs 2 Pairs Timer 1 - 999 minutes with alarm, continuous Safety Device...
Section 4 Installation Instructions Mini Pro 300V Power Supply is actually a pre-installed instrument. As long as it is placed on a sturdy and level surface in a safe, dry place, and further connects with well-prepared electrophoresis system, it is ready for operation.
Section 5 Operation Instructions 5.1 Control interface You will be able to find six buttons and 4 LED indicators from the faceplate. The LED indicates the status of the unit. a. Setup Mode (before pressing RUN/Start) LED light indicates your current parameter. If you are setting a value for Voltage, the Voltage LED will be lit.
– Start or stop the unit 5.2 Start the operation Note: To operate under constant voltage or constant current modes, adjust the other parameter to the maximum value. For example, to operate under constant voltage, adjust current to max before running using constant voltage, and vice versa.
Page 15
Timer button, and then press key or key to adjust timer accordingly. Note: When timer is set as “0”, it means continuous operation. 9. Press Key to start the run. Once the electrophoresis starts, the LED light next to the button will light up.
Check AC power cord connections at both connected ends. Use the correct cords. The fuse has blown Replace the fuse Repeated fuse Contact Major Science service Hardware failure broken department Electrophoresis leads Check the connections to the power are not connected to...
Page 17
2. Call Technical Service or e-mail to service@majorsci.com 3. If the unit must be shipped back for repair, contact Major Science or the distributor for a Return Authorization Number and shipping instructions. The unit will be repaired and returned to you as quickly as possible.
Page 18
5. Snap the cover closed. Maintenance Mini Pro 300V Power Supply uses all solid-state components and should require no maintenance or recalibration under normal use. If the unit must be returned for repair, contact our SERVICE DEPARTMENT or your local...
Section 7 Ordering Information Cat. No. Description MINI-300 Mini Pro 300V Power Supply Single Electrophoresis Package MJS-Mini300 Package of MJ-105-S and Mini-300 MJR-Mini300 Package of MJ-105-R and Mini-300 MT-Mini300 Package of MT-108 and Mini-300 Dual Electrophoresis Package MJSS-Mini300 Package of 2 x MJ-105-S and Mini300...