SprintScan 35LE Service Manual
Checking for Main PC Board Failure
1. Check the output signal at test point TP2. Figure 3-6 shows the correct output
waveform.
Figure 3-6. Post-processing CCD Output Waveform at TP2
2. If the output waveform is incorrect, replace the main PC board.
Note: Its level must be even to level GND.
Checking SCSI Port
If the READY LED comes on but the scanner does not respond to a Start Scanning
command from the host computer:
1. Check to see if the SCSI ID is correct for both the software program and the
scanner.
2. Set the SCSI ID number to 7 and power up the scanner to enter the self-test mode.
(Shortly after, the scanner carriage will traverse up and down at a slow speed.)
Note: If the scanner fails to perform this routine, a main PC board malfunction is
detected.
3. Power off and set the SCSI ID to its original settings.
Note: When the SCSI bus is confirmed OK, replace the main PC board.
Votage
~
0 3.5V
~
2.5 msec
64 msec
28
Troubleshooting
Time