ReliOn Confirm User Instruction Manual page 49

Hide thumbs Also See for Confirm:
Table of Contents

Advertisement

Available languages

Available languages

It is Recommended to Perform a Control Solution Test:
• Whenever you suspect the meter or test strips may not be
functioning properly.
• If test results appear to be abnormally high or low or are not
consistent with clinical symptoms.
• The test strip bottle has been left open or has been exposed to
temperatures below 34°F (1°C) or above 86°F (30°C), or humidity
levels above 80%.
• To check your technique.
• When the ReliOn
dropped or stored below 32°F (0°C) or above 122°F (50°C).
• When you open a new bottle of test strips.
Important:
• Use only with ReliOn
ReliOn
Confirm/Micro blood glucose test strips.
®
Confirm blood glucose meter has been
®
Confirm blood glucose meter and
®
49

Hide quick links:

Advertisement

Chapters

Table of Contents
loading

Table of Contents