Safety System:
Width:
Height:
Depth:
Weight:
Shipping Weight:
Operating Ambient
Temperature:
Exhaust:
Oxygen Output Temperature: Less than Ambient + 6°F (+3°C)
Cabinet:
Regulatory Listing:
Electrical:
Placement:
Tubing:
Relative Humidity:
Time of Operation:
Recommended Storage and
Shipping Temperature:
NOTE: Stated concentration levels achieved after initial warm-up
period (approximately 30 minutes).
Part No 1143482
SECTION 4—TYPICAL PRODUCT PARAMETERS
Current overload or line surge shutdown. High
temperature compressor shutdown. High
Pressure Alarm w/compressor shutdown. Low
Pressure Alarm w/compressor shutdown.
Battery Free Power Loss Alarm. SensO
Oxygen System Possible Obstruction Alert.
15 in ± 3/8 in (38.1 cm ± 1 cm)
23 in ± 3/8 in (58.4 cm ± 1 cm)
12 in ± 3/8 in (30.5 cm ± 1 cm)
43 lbs ± 2 lbs (19.5 kg ± 1 kg) Perfecto
45 lbs ± 2 lbs (20.5 kg ± 1 kg) Perfecto
Models
40 lbs ± 2 lbs (18.1 kg ± 1 kg) Perfecto
48 lbs ± 2 lbs (21.8 kg ± 1 kg) Perfecto
50 lbs ± 2 lbs (22.8 kg ± 1 kg) Perfecto
Models
45 lbs ± 2 lbs (20.4 kg ± 1 kg) Perfecto
50°F - 95°F (10°C - 35°C) at 20-60% relative
humidity Perfecto
Model
50°F - 90°F (10°C - 32°C) at 20-60% relative
humidity Perfecto
Less than Ambient + 35°F (+19°C)
Impact Resistant flame-retardant plastic cabinet
that conforms to UL94V-0.
Double Insulated Product ETL certified
complying with UL1431 and UL1097 ETL
certified to CSA C22.2 No. 68.
No extension cords.
No closer than twelve inches from any wall,
furniture, draperies, or similar surfaces.
7 ft cannula with a maximum 50 ft of
Crush-Proof Tubing (DO NOT pinch).
20 to 60%
Up to 24 hours per day.
-20°F to 150°F (-29°C to 65°C), at 15-95%
relative humidity.