MESSAGE
c 3 . 8
(example)
contrl
PROBLEM
3) The Test Strip was
inserted before insert
strip appeared on the
display.
4) The center of the Test Strip
Holder where the small hole
is located is raised.
If this message appears after
a blood test, your blood
sample was too small,
smeared, or another drop was
added after the test began.
WHAT TO DO
Repeat the test. Wait for
insert strip before
inserting a new Test Strip.
Depress the center by
pressing forward on the
base, or by pressing down
on the center, or by inserting
a check strip into the Test
Strip Holder.
Repeat the test with a new
Test Strip and a larger drop
of blood.
45