Bar Code Calibration Procedure Without Remote Display - Metrologic MS2020 Stratos Installation And User Manual

Metrologic ms2020 stratos series barcode scanner
Hide thumbs Also See for MS2020 Stratos:
Table of Contents

Advertisement

S
O
CALE
PERATION
C
ALIBRATION

Bar Code Calibration Procedure without Remote Display

The following calibration procedure can be used when the remote scale display is
not present. This procedure requires that the scanner/scale have a software
serial number of 15001, or greater. The beeper volume switch is used to
advance to the next stage of calibration and the LED display notifies the operator
which 'calibration stage' (1 through 5) is active.
1.
Temporarily remove the platter and place it in a safe location.
It is the responsibility of the owner of the scale to confirm compliance
with the relevant Weights and Measures statutes and regulations
applicable in your area by checking with the appropriate government
agency before placing a newly calibrated unit into service or
removing any official seals.
2.
If this is a currently installed scanner/scale in need of calibration, cut and
remove the calibration switch cover seal. If this is a new installation, cut and
remove the factory-applied adhesive seal.
Follow all Electo-Static Discharge (ESD) procedures when exposing
internal scanner/scale components.
3.
Remove the M3 screw securing the calibration switch/button cover. Place
the cover and screw in a safe location.
76
Figure 69. Calibration Switch Plate Cover Removal

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents