Scale Calibration Procedure (lbs. or kg) with Remote Display
12. Remove the remaining half load from the scale then press clear (C). The
message done will flash briefly on the display.
13. Exit calibration mode. Press and hold the test button for at least 3
seconds then release. With no load on the scale the display should read
0.00 lb. or 0.000 kg.
14. The unit's calibration must now be verified as required by state and/or local
Weights and Measures regulations (starting on page 80).
Need to Start Over?
If for any reason you need exit the calibration mode or restart the process press
the test button then the clear button.