Technical Data - Miele PG 8527 Operating Instructions Manual

Washer-disinfector
Table of Contents

Advertisement

Height
Width
Depth
Weight (net)
Operating weight
Voltage
Connected load
Fuse rating
Compressed air connection
Steam connection:
with electric TA
with steam TA
Water pressure (flow rate)
Water connection:
Cold, hot and de-ionised (AD) water
Ambient temperature
Relative humidity: maximum
decreasing proportionately to
Altitude
Soiling level
(according to IEC/EN 61010-1)
Protection category
(according to IEC 60529)
Noise level in dB (A), Sound level LpA
in washing and drying
Test marks
! mark
Manufacturer's address
* if the steam pressure lies between 250 - 300 kPa (2.5 – 3.0 bar) the steam drying
unit (TA) will only reach a max. drying temperature of 90 - 100 °C.
§
for AD water connection with special version with DK+WRG max. 20°C
#
If installed at altitudes above 1500 m the boiling point of the suds solution will be
lower. Disinfecting temperature parameters should be lowered and the holding
time increased (A
value).
0
250 cm (minimum room height)
115 cm
87 cm
Approx. 550 kg
Approx. 800 kg
see data plate
see data plate
see data plate
600 kPa (6.0 bar) required for steam
operation
250 - 1000 kPa / 140 - 180 °C
250 - 600 kPa * / 140 - 164 °C
200 - 1000 kPa overpressure
up to max. 70 °C
§
5 °C to 40 °C
80 % for temperatures to 31 °C
50 % for temperatures to 40 °C
Max. 1500 m
P2
IP20 (Dust permeation)
< 70
VDE, radio interference suppression
MDD-Guidelines 93/42/EWG, Class IIb
Miele & Cie. KG, Carl-Miele-Str. 29,
33332 Gütersloh, Germany

Technical data

#