This machine should be
connected to the mains water
supply and to the drainage system
in accordance with local and
national water regulations. Please
refer to the installation diagram
supplied with the machine.
– If the water supply has a high iron
content there is a danger of
corrosion occurring on items being
reprocessed in the machine, as well
as in the machine itself.
If the chloride content of the water
exceeds 100 mg/l the risk of
corrosion to items being reprocessed
in the machine will be further
increased.
– If there are too many impurities in the
local water supply, it will be
necessary to operate the steam
condenser with de-mineralised
water.
– The machine is supplied as standard
for connection to a cold, hot and AD
water supply. Where hot water
connection is not available on-site,
the hot water connection inlet valve
must be connected to the cold water
supply.
This allows the two valves to be
controlled in parallel in the
inlet
programme stage. Programme
duration will be shortened.