Response Message Data Types; Status Reporting - Escort 3136A Operation Manual

50000 count multimeter
Table of Contents

Advertisement

CONFigure is a root level keyword with the second level keyword, VOLTage,
and 0.5 is the command parameter. The query command ends with a question
mark "?".
Note:
SCPI stems from IEEE488.1 and IEEE 488.2. Although the IEEE 488.2 standard addressed
some instrument measurements, it principally dealt with common commands and syntax or data
formats. Please refer to the IEEE488.2 and SCPI reference manual for more information.
7-4-2 Data Types of Responding Message
Responding message is the data from the meter in response to a query. A query
is a command followed by a question mark. Table 7-2 is explanation for data
types.
Table 7-2. Data Types of Responding Message
Data Type
<NR1>
An integer
This numeric representation has an explicit radix
<NR2>
point.
This representation has an explicit radix point and
<NR3>
an exponent.
<Numeric value> Accepts <NR1>, <NR2> and <NR3> data types.
Flexible numeric representation
<NRf>
(Only positive integers).
Single ASCII-encoded byte, is return for the settings
<Boolean>
query.
ASCII-encoded bytes corresponding to the short
<Literal>
form of the literal used as the command parameter.

7-4-3 Status Reporting

The meter status registers conform to the SCPI and IEEE-488.2 standards.
Service Requests (IEEE-488 Only and Status Registers)
Service requests let a meter on the IEEE-488 bus get the attention of the host.
Service requests are sent over the service request (SRQ) bus line.
If more than one instrument on the bus is capable of sending service requests,
the host can decide which instrument made the request by taking a "serial poll."
Each instrument on the bus responds to the poll by sending the contents of its Status
Byte Register.
If an instrument on the bus has made a service request, the request service bit
(RQS, bit 6) of its Status Byte Register will be set to 1, identifying it as an
Explanation
73
Example
+10000, -10000,
123, -100
+13.234, -.00002,
3.4567
+1.2E+2, +0.1E+0,
-0.12E-3, 9E+9
100, 255, 16
0 or 1,OFF or ON