Querying Parameter Settings; Scpi Command Terminators; Ieee-488.2 Common Commands - Agilent Technologies E3633A User Manual

Dc power supplies
Hide thumbs Also See for E3633A:
Table of Contents

Advertisement

Chapter 4 Remote Interface Reference
An Introduction to the SCPI Language

Querying Parameter Settings

You can query the value of most parameters by adding a question mark (?) to the
command. For example, the following command sets the output current to 5 amps:
''CURR 5''
You can query the value by executing:
''CURR?''
You can also query the minimum or maximum value allowed with the present function
as follows:
''CURR? MAX''
''CURR? MIN''
C a u t i o n
If you send two query commands without reading the response from the first, and then
attempt to read the second response, you may receive some data from the first response
followed by the complete second response. To avoid this, do not send a query
command without reading the response. When you cannot avoid this situation, send
a device clear before sending the second query command.

SCPI Command Terminators

A command string sent to the power supply must terminate with a <new line>
character. The IEEE-488 EOI (end-or-identify) message is interpreted as a <new line>
character and can be used to terminate a command string in place of a <new line>
character. A <carriage return> followed by a <new line> is also accepted. Command
string termination will always reset the current SCPI command path to the root level.
The <new line> character has the ASCII decimal code of 10.

IEEE-488.2 Common Commands

The IEEE-488.2 standard defines a set of common commands that perform functions
like reset, self-test, and status operations. Common commands always begin with an
asterisk ( * ), are four to five characters in length, and may include one or more
parameters. The command keyword is separated from the first parameter by a blank
space. Use a semicolon ( ; ) to separate multiple commands as shown below:
104
''*RST; *CLS; *ESE 32; *OPC?"