Chapter 3 Front-Panel Operation
Remote Interface Configuration
HP-IB Address
Each device on the HP-IB (IEEE-488) interface must have a unique address. You can
set the power supply's address to any value between 0 and 30. The current address is
displayed momentarily on the front panel when you turn on the power supply. The
address is set to "05" when the power supply is shipped from the factory.
The HP-IB address can be set from the front-panel only.
• The address is stored in non-volatile memory, and does not change when power
has been off or after a remote interface reset.
• Your HP-IB bus controller has its own address. Be sure to avoid the bus controller's
address for any instrument on the interface bus. Hewlett-Packard controllers
generally use address "21".

Baud Rate Selection (RS-232)

You can select one of six baud rates for RS-232 operation. The rate is set to 9600 baud
when the power supply is shipped from the factory.
The baud rate can be set from the front-panel only.
• Select one of the following: 300, 600, 1200, 2400, 4800, 9600 baud. The factory
setting is 9600 baud.
• The baud rate selection is stored in non-volatile memory, and does not change
when power has been off or after a remote interface reset.

Parity Selection (RS-232)

You can select the parity for RS-232 operation. The power supply is configured for
no parity and 8 data bits when shipped from the factory.
The parity can be set from the front-panel only.
• Select one of the following: None (8 data bits, factory setting), Even (7 data bits),
or Odd (7 data bits). When you set the parity, you are indirectly setting the number
of data bits.
• The parity selection is stored in non-volatile memory, and does not change when
power has been off or after a remote interface reset.
3