Attaching a Stage
(1) Turn the coarse focus handle to remove the
cushioning material from the substage
section.
(2) Turn the coarse focus handle until the
elevating section is brought to the lowermost
position.
(3) Place the stage on the substage and fix into
place with the tool stored in the back of the
microscope.
(4) Place a specimen holder on the stage and
tighten screws.
5
Attaching a Condenser
(1) Turn the coarse focus handle until the
elevating section is raised to the uppermost
position.
(2) Turn the condenser focus knob until the
substage is brought to the lowermost
position.
(3) Insert a condenser and adjust so that it faces
toward the front. Secure in place with the tool
stored in the back of the microscope.
(4) Turn the condenser focus knob until the
substage is raised to the uppermost position.