Note
When disassembling, take care of small parts such as screws and washers. Put all small parts
down so that they can be reassembled in the right sequence.
Remove legrests.
Release seat (1) and tilt forwards.
Disconnect the remote bus cable from
the electronics module or ACT.
Lift the seat out of the front anchorage
and remove it. When doing so, get the
help of a second person because the
seat is heavy.
Remove both battery cases as described
in Chapter 7.7.1.
Disconnect the motor plug from the
motor to be replaced from the electronics
module.
The motor cable is secured inside the
frames with cable ties. Remove the cable
ties with the oblique pliers.
Place the vehicle upside down so that
the wheels are facing upwards. When
doing so, get the help of a second
person because the vehicle is heavy.
Heat countersunk screw TX40 (1) with
hairdryer to release screw blocker
system.
Remove the torx screw (1) with a TX40
spanner.
Pull the complete wheel off the wheel
hub.
A washer can be used to compensate the
tolerance between wheel hub and shaft
shoulder.
When replacing the countersunk screw
TX40 (1) always use a high-strength
screw blocker system such as Loctite
270!