SERVICE MANUAL
Component
Drive units,
coupling
mechanism
Legrests
Power legrests
(if fitted)
Lighting (if
fitted)
Batteries
Battery case
Remote /
electronics
module
Drive program
26
Check
 Check functions in
drive and push
modes
 Check coupling
mechanism
 Check welded
seams, interlocking,
screws, footplates
 Check cable
 Check contacts
 check functions
 Check cable
 Check function
 Check batteries for
damage
 Check battery
voltage
 Check contacts and
terminals
 Check locking
system, it must
engage completely.
 Remote, status
display blinking
 Fixings
 Cables and
connecting plugs
 Drive lever function
 Power supply
 Check drive
electronics program
version
Remedy
 Replace motor if
necessary.
 Tighten
screws/nuts,
adjust or replace if
necessary
 Tighten, replace if
necessary
 Replace cable if
necessary
 Replace lamp or
cable if necessary
 Replace batteries
if necessary
 Charge batteries
 Clean contacts
and terminals
 Replace if
necessary
 Evaluate
error/blink code
 Tighten fixings,
replace if
necessary
 Tighten cables
and connecting
plugs, replace if
necessary
 Replace drive
lever if necessary
 Replace remote if
necessary
 Tighten cables
and connecting
plugs, replace if
necessary
 Update software if
newer version
available.
®
Invacare
- Bora
Notes
See chapter 7.7.
See operating
manual
Please refer to the
safety information
in Chapter 7.7 for
handling batteries
See chapter 7.6.