Adjustment Settings - T&D TR-5xU Series User Manual

T&d recorder for windows
Table of Contents

Advertisement

Adjustment Settings:from the [Settings] Menu
By setting adjustment values beforehand, you can record and display
the post- adjusted measurement values. You can choose from two
adjustment methods: 1-point and 2-point. Adjustment will be carried out
using an adjustment equation of Y = aX + b; where X is the pre-
adjusted measurement value and Y is the post-adjusted value.
Rough Guidelines for Adjustment
1 point Adjustment: Use when measuring in a range of ± 20℃ .
2 point Adjustment: Use when measuring in a wide range.
NOTE:
- When adjusting at 2 points, make sure the difference in points is at least 10℃ .
- When measuring in a wide range and adjusting at 2 points, make sure that the adjustment
values reflect the range you are measuring.
- We cannot guarantee that after carrying out adjustment the measuring accuracy will improve for
all measuring ranges.
1.
Check the channel you wish to adjust.
Check here
Basic Function - TR-71U/72U/73U
29
[Save File] Button
[Read File] Button
[History] Column