Disassembly; Performance Tests - Fluke 15B+ Calibration Manual

Hide thumbs Also See for 15B+:
Table of Contents

Advertisement

15B+/17B+/18B+
Calibration Manual

Disassembly

To prevent false readings, which could lead to possible electric
shock or personal injury, replace the batteries as soon as the
low battery indicator () appears.
To prevent damage or injury:
To remove the battery door and battery for adjustments and calibration, see
Figure 1.

Performance Tests

Use the performance tests to make sure the UUT is operating properly and to
make sure the UUT is accurate. If the UUT fails any part of the performance test,
repair and/or calibration adjustment are required. See the Calibration sections for
calibration adjustment. If the UUT still fails to meet the range indicated, see How
to Contact Fluke.
Table 2 lists the equipment required for the performance tests.
Equipment
Calibrator
10
XWWarning
Install ONLY replacement fuses with the specified
amperage, voltage, and interrupt ratings.
Disconnect test leads before opening the case or the
battery door.
Figure 1. Remove the battery
Table 2. Required Equipment for Performance Test 15B+ and 17B+
FLUKE 5522A
Recommended Model
icr01.eps

Advertisement

Table of Contents
loading

This manual is also suitable for:

17b+18b+

Table of Contents