Invacare ProntoM41 Service Manual page 19

With surestep
Table of Contents

Advertisement

®
Invacare
- PRONTO M41
Number
of Flashes
8
9
10
11
Diagnostics
Error Code
Code
Description
E08
Controller
Fault
E09
Communicati
ons Fault
General
Fault
Incompatible/
incorrect
Remote
Sub
Details of
Code*
Error Code
00
Controller fault
01
RAM fault
02
ROM fault
03
CPU fault
04
EEPROM fault
05
Watchdog fault
06
Stack fault
07
Software fault
08
Power-up testing
fault
09
Relay fault or
precharge fault
10
Bridge fault or
disable all fault
11
Electronics fault:
Thermistor
12
Calibration setting
fault
00
Remote connection
lost
01
Low communication
mode
00
General fault
00
Incompatible/
incorrect Remote
SERVICE MANUAL
Possible
Solution
Check connections
and wiring. Replace
power module.
Check connections
and wiring. Replace
Bus cable.
Check all
connections and
wiring. Contact
Invacare Technical
Service.
Wrong type of
remote connected.
Ensure the branding
of the joystick
matches that of
controller unit.
19