PROCEDURE 7
ADJUSTING FOOTREST
Position.
1. Loosen the hex screws on the mounting bracket.
2. Slide the footrest assembly along the frame to the desired position.
3. Tighten the hex screws on the mounting bracket.
Height.
1. Loosen the set screw securing the upper and lower footrest support tubes.
2. Reposition the footrest up or down as desired.
3. Tighten the set screw to secure the upper and lower footrest support tubes.
Angle.
1. Loosen the locknuts securing the footplate mounting bracket.
2. Adjust the angle of the footplate as desired.
3. Tighten the locknuts securing the footplate mounting bracket.
Depth.
1. Remove the locknuts and button screws securing the footplate mounting bracket.
2. Reposition the footplate to the desired set of mounting holes.
3. Install the button screws through the footplate, footstrap and footplate mounting bracket.
4. Secure the button screws with the locknuts. Tighten securely.
HEIGHT ADJUSTMENT
Upper Support
Tube
FIGURE 3 - ADJUSTING THE FIXED FOOTREST
T-3 TENNIS ELITE INDIVIDUAL FOOTPLATES -
BEFORE 4/2000 (FIGURES 2 AND 3)
The position of the footrest, camber tube, seat height, the tautness of the
back upholstery, as well as the user's condition are directly related to the
chair's stability. Any change to one (1) or any combination of the five (5) may
cause the chair to decrease in stability. Use EXTREME caution when using a
new seating position.
T-3™ and T-4™ Series
Set Screw
Mounting Holes
WARNING
52
FOOTREST
DEPTH ADJUSTMENT
Footplate
Locknuts
Part No. 1054916
Button
Screws