PROCEDURE 6
QUICK RELEASE LEVERS - Make sure the quick-release levers are pointing
towards the rear of the wheelchair and are in the CLOSED position BEFORE
using the wheelchair, otherwise personal injury or damage to the wheelchair
may result.
STANDARD - Make sure the hex screws and locknuts are securely tightened
BEFORE using the wheelchair, otherwise personal injury or damage to the
wheelchair may result.
9. Close the camber clamp. Refer to OPENING/CLOSING CLAMPS in this procedure of the
manual.
10. Repeat STEPS 1-9 for the opposite side of the wheelchair.
Make sure the detent pin and locking pins of the quick/quad-release axle are
fully released BEFORE operating the wheelchair.
The locking pins MUST be protruding past the inside of the rear wheel axle
bushing for a positive lock.
Keep locking pins clean.
11. If the locking pins of the quick/quad-release axles are not protruding past the inside of the
axle bushing or there is too much movement of the rear wheel assembly in a back and
forth position, refer to ADJUSTING QUICK-RELEASE AXLES or REMOVING THE PLAY
FROM THE REAR WHEELS in this procedure of the manual.
12. Refer to the following chart to determine if it will be necessary to adjust the toe in/toe out
of the wheelchair.
0 and 9, 0 and 12, 3 and 9
3 and 12, 6 and 9, 6 and 12
* Refer to ADJUSTING TOE IN/TOE OUT in this procedure of the manual.
Indexing Marks
FIGURE 12 - REPOSITIONING CAMBER INSERTS
(ADJUSTING REAR WHEEL CAMBER)
T-3™ and T-4™ Series
WARNING
WARNING
Camber Insert
Degree Range
0 and 3, 0 and 6,
3 and 6, 9 and 12
Camber Insert - Rotate 180
Indexing Ring
42
Adjustment
Required
NO
* YES
Rear View of Wheelchair
o
Frame Bracket
WHEELS
Axle Tube
Part No. 1054916