Installing the Reliant Scale

NOTE: For this procedure, refer to FIGURE 8.2.
Patient and sling MUST be removed from the lift during ALL installation proce-
dures.
1. Position the load cell assembly of the Reliant Scale into the boom mounting bracket. 
Refer to Detail ʺAʺ in FIGURE 8.2.
NOTE: Use 1/4‐inch nylon washer. 
2. Secure the Reliant Scale to the boom mounting bracket with a shoulder bolt, two nylon 
washers and a locknut. Securely tighten. Refer to Detail ʺAʺ in FIGURE 8.2 for washer 
orientation.
NOTE: Ensure the shaft of the shoulder bolt passes through both sides of the boom mounting 
bracket.
3. Insert swivel bar pin with two nylon washers through the swivel bar. Refer to Detail 
ʺBʺ in FIGURE 8.2.
4. Align the mounting holes in the swivel bar pin with the mounting holes in the load 
cell assembly. Refer to Detail ʺBʺ in FIGURE 8.2.
5. Secure swivel bar pin to the load cell assembly with the provided mounting screw and 
locknut. Securely tighten. Refer to Detail ʺBʺ in FIGURE 8.2.
After ANY adjustments, repair or service and BEFORE use, make sure all attaching
hardware is tightened securely - otherwise injury or damage may occur.
6. Weigh the patient. Refer to Operating the Scale on page 34 and Weighing the Patient 
on page 35.
DETAIL "A"
Boom
Locknut
Boom
Mounting
Bracket
Nylon
Washer
Reliant
Scale
Part No 1078988
WARNING
WARNING
Reliant
Scale
Mounting
Shoulder
Bolt
Nylon
Nylon
Washer
Washers
Load Cell
Assembly
FIGURE 8.2 Installing the Reliant Scale