SECTION 7—MAINTENANCE
Checking and Tightening Mast Pivot Bolt
NOTE: For this procedure, refer to FIGURE 7.3.
1. Lift up the back of the rubber boot and slide it off the mast along the boom.
2. Check that the bolt is through the bracket and the locknut is tight and secure.
3. If needed, do one or more of the following:
• tighten locknut and back‐off the locknut 1/8 of a turn.
• replace the locknut.
4. Reposition the rubber boot.
FIGURE 7.3 Checking and Tightening Mast Pivot Bolt
Replacing the Swivel Bar
After the first year of use, the hooks of the swivel bar and mounting brackets of the
boom should be inspected every six months to determine the extent of wear. If
these parts become worn, replacement must be made.
NOTE: For this procedure, refer to FIGURE 7.4.
1. The swivel bar comes attached to the boom with bolt, nut, washers, and pinch guard.
NOTE: Washers should be placed against both sides of the boom mounting bracket to eliminate
damage.
NOTE: The swivel bar has three hooks on both ends to attach the sling that supports the patient
during lift.
2. Remove existing hardware and replace the swivel bar.
DO NOT overtighten the nut and bolt. This damages the mounting bracket.
Electric Portable Patient Lift - RPA600-1E
Washer
Bolt
Mast
WARNING
CAUTION
30
Rubber Boot
Boom
Spacer
(if equipped)
Locknut
Bracket
Part No 1078988