Daily/After each use:
Disassemble the mask components according to the
disassembly instructions.
Thoroughly clean the separated mask components (excluding
headgear), by gently rubbing in warm (approximately 86°F/30°C)
water using mild soap.
To optimize the mask seal, facial oils should be removed from
the cushion after use.
Use a soft bristle brush to clean the vent.
Inspect each component and, if required, repeat washing until
visually clean.
Rinse all components well with drinking quality water and allow
to air dry out of direct sunlight.
When all components are dry, reassemble according to the
reassembly instructions.
Weekly:
Handwash the headgear.
Reprocessing the mask between patients
This mask should be reprocessed when used between patients.
Cleaning, disinfection and sterilization instructions are available from
the ResMed website,
www.resmed.com/masks/sterilization/americas. If you do not have
internet access, please contact your ResMed representative.