Problem
Display error message:
Exxxx
Call Service
(where xxxx defines
an error)
or
MOTOR FAULT
Call Service
or
UPLOAD LANGUAGES
Call Service
or
DATA LOST
Call Service
The following message
is displayed on the
LCD after you try to
update settings using
the Data Card:
Card Error
Remove Card
The following message
is displayed on the
LCD after you try to
copy data to a Data
Card:
Card Error
Remove Card
The following message
is displayed on the
LCD after you try to
update the settings
using the Data Card:
Settings Invalid
Remove Card
Possible Cause
Component failure.
The Data Card is not inserted
correctly.
The Data Card is only partly
inserted.
You may have removed the Data
Card before settings were copied
to the S8 Escape.
The card is not a ResScan Data
Card.
The Data Card is not inserted
correctly.
The card is not a ResScan Data
Card.
The identification details on the
Data Card do not match the
details on your device.
Solution
Call Service.
Ensure that the Data Card is
inserted with the arrow facing up
and pointing towards the device.
Ensure that the device is switched
on and in standby mode before
you insert the Data Card.
Ensure that the Data Card is
inserted into the slot on the Data
Card module as far as it will go.
Reinsert the Data Card and wait
for the Settings Success
Remove Card message to
appear on the LCD.
Only use a ResScan Data Card.
Ensure that the Data Card is
inserted with the arrow facing up
and pointing toward the device.
Use a ResScan Data Card, as
specified in this manual.
Contact your clinician
immediately.
T
ROUBLESHOOTING
27