Scpi Command Reference; Abort - HP E1366A User Manual

Hewlett-packard rf multiplexers user's manual
Table of Contents

Advertisement

SCPI Command Reference

ABORt

Subsystem Syntax
Comments
Example
40
HP E1366A/E1367A RF Multiplexers Command Reference
This section describes the Standard Commands for Programmable
Instruments (SCPI) commands for the RF multiplexers. Commands are
listed alphabetically by subsystem and within each subsystem.
The
ABORt
subsystem stops a scan in progress when the scan is enabled via
the interface, and the trigger modes are
TRIGger:SOURce HOLD
ABORt
ABORt
operation invalidates the current channel list and sets
ARM:COUNt 1
(one scanning cycle per
INITiate CONTinuous OFF
TRIGger:SOURce IMMediate
Stopping Scans Enabled from Interface: When a scan is enabled
from the interface, use an interface
HP E1301A front panel " Reset Instr" or "Clear Instr" key to stop
the scan.
When the scan is enabled from the interface and the trigger source is
TRIGger:SOURce BUS
the HP E1301A front panel " Reset Instr" or "Clear Instr" key to
stop the scan.
Stopping Scans Enabled from Front Panel: When a scan is
enabled from the HP E1301A front panel, execute
the interface or the front panel " Reset Instr" or " Clear Instr" key to
stop the scan.
Related Commands:
TRIGger
Stopping a Scan with
ABORt
TRIG:SOUR BUS
INIT:CONT ON
SCAN (@100:103)
INIT
.
.
ABOR
TRIGger:SOURce BUS
.
INITiate
(no continuous scanning cycles), and sets
(continuous internal triggering).
CLEAR
or
TRIGger:SOURce HOLD
,
ARM
INITiate:CONTinuous
!
*TRG
! Set continuous scanning
! Scan channels 00-03
! Start scan, close channel 00
! Abort scan in progress
or
command), sets
command or the
, use
ABORt
or
*RST
entered via
,
[ROUTe:]SCAN
command is trigger source
Chapter 5
,

Advertisement

Table of Contents
loading

This manual is also suitable for:

E1367a

Table of Contents