Troubleshooting Error Codes; Product Specifications - HoMedics TT-200 Instruction Manual

Deluxe instant ear thermometer
Table of Contents

Advertisement

Available languages

Available languages

TROUBLESHOOTING/ERROR CODES
Condition
Possible Cause Corrective Action
Consistent low
Temperature
temperature
reading is below
readings ("Lo")
measurement range
(below 50˚F/10°C).
The probe is not
positioned properly
in ear.
The waterproof
probe
is dirty.
Consistent high
Temperature
temperature
reading is above
readings ("Hi")
measurement range
(above 122˚F/50°C).
Consistent error
Operating
messages ("Err")
temperature
is not within the
permissible range
for measurement.
Low battery warning Battery power
is low.
Low battery warning
Battery power is
with "Lo" symbol
too low for the
thermometer to
function.
Product Design-
Measurement range is
50˚F (10°C) to 122˚F
(50°C). Contact your
physician if you have or
suspect any medical
problems.
Make sure probe is
properly positioned
against the opening
of the ear canal
Clean the probe with a
cotton swab moistened
with alcohol.
Product Design-
Measurement range is
50˚F (10°C) to 122˚F
(50°C). Contact your
physician if you have or
suspect any medical
problems.
Product Design-
Operating temperature is
60.8˚F (16°C) to 104˚F
(40°C).
Replace battery (see
"Inserting and Replacing
Batteries").
Replace battery
immediately (see "Inserting
and Replacing Batteries").
13

PRODUCT SPECIFICATIONS

Model Numbers
Power Source
Battery Life
Sensor
Temperature Indication
Accuracy
Measurement Range
Weight
Operating Temperature
Operating Humidity
Storage Temperature
Storage Humidity
Contents
NOTE: These specifications are subject to change without notice. This ear
thermometer meets requirements established in ASTM Standard E1965-98.
Model Numbers TT-200/TT-201
DC 3 Volt (lithium battery CR2032)
1,000 Measurements in day mode or
300 measurements in night mode.
Thermopile
In 4 digits, every 0.1˚F or ˚C
+/- 0.4˚F/0.2°C between 96.8˚F -
102.2˚F (36.0°C - 39.0°C)
+/- 0.5˚F/0.3°C less than 96.8˚F
(36.0°C) or greater than 102.2˚F
(39.0°C)
50˚F - 122˚F (10˚C - 50˚C)
0.08 lbs (with battery)
60.8˚F - 104˚F (16˚C - 40˚C)
Relative humidity up to 95%
(non-condensing)
-4˚F - 122˚F (-20˚C - 50˚C)
Relative humidity up to 95%
(non-condensing)
Ear Thermometer & Battery, Storage Case,
Instruction Manual