Storage
Temperature
Vibration
Weight
Dimensions
Power
Battery Life

KTC Specifications

Only approved accessories can be used with the Product.
For details, please refer to the separate Safety
Instructions.
Measurement Range
Accuracy
Cable Length
-20 °C to 60 °C (-4 °F to 149 °F)
2.5 G, IEC 68-2-6
0.322 kg (0.7099 lb)
17.69 cm (6.965 in) H x 16.36 cm (6.441 in) L
x 5.18 cm (2.039 in) W
3 AAA /LR03 type-approved batteries.
(For a list of type-approved batteries, please refer to
the separate Safety Instructions.)
4 hours with laser and backlight on; 100 hours
with laser and backlight off, at 100 % duty cycle
(Product continuously on)
Note
-40 °C to 260 °C (-40 °F to 500 °F)
±1.1 °C (±2.0 °F) from 0 °C to 260 °C (32 °F to
500 °F). Typically within 1.1 °C (2.0 °F) from -
40 °C to 0 °C (-40 °F to 32 °F)
1 m (40 in) KTC cable with standard
miniature thermocouple connector and bead
termination
Infrared Thermometer
Specifications