Ventilation Requirements

5000 and 6000A Series Ventilation Requirements

6000L Series Ventilation Requirements

7000 Series Ventilation Requirements

InfiniiVision Oscilloscopes User's Guide
The air intake and exhaust areas must be free from obstructions. Unrestricted air flow is
required for proper cooling. Always ensure that the air intake and exhaust areas are free
from obstructions.
The fan draws air in from underneath the oscilloscope and pushes it out behind the
oscilloscope.
When using the oscilloscope in a bench-top setting, provide at least 4" (100 mm)
clearance behind and above the oscilloscope for proper cooling.
The fan draws air from the left and pushes it out to the right. Ensure the rack in which the
oscilloscope is mounted is properly ventilated.
The fan draws air in from the sides of the oscilloscope and pushes it out behind the
oscilloscope.
When using the oscilloscope in a bench-top setting, provide at least 2" clearance at the
sides and 4" (100 mm) clearance above and behind the oscilloscope for proper cooling.
2
Initial Setup
47