11.3. Calibration-Related Functions

Function No.
Parameter or
Setting range
Calf01
0
4
to
Calf02
0
5
to
Calf03
1
6
to
Calf04
1
999999
to
Calf05
0
100
to
Calf06
0. 0
5. 0
to
Calf07
0. 0
9. 9
to
Calf08
0. 0
9. 9
to
Calf09
0
9
to
Calf10
0
1
to
Calf11
0
1
to
Calf12
0
1
to
Description
Unit of the weight value
0: None
1: g
2: kg
3: t
4: lb (USA version)
Decimal point position of the weight value: 10
Minimum division (d) of the weight value
1: 1
2: 2
3: 5
4: 10
5: 20
6: 50
Weighing capacity of the weighing instrument
The decimal point position is the same as
Calf02
the setting of
Range to enable zero adjustment by the
ZERO key
Expressed in percent of the weighing
capacity with the calibration zero point as
the center
Zero tracking time (Unit: second)
When 0.0, zero tracking will not be performed.
Zero tracking band (Unit: 0.1 d)
When 0.0, zero tracking will not be performed.
Stability detection time (Unit: second)
When 0.0, stability detection will not be
performed.
Stability detection band (Unit: d)
When 0, stability detection will not be
performed.
Tare and zero adjustment when the weight
value is unstable.
0: Disables both functions.
1: Enables both functions.
Tare when the gross weight is negative.
0: Disables tare.
1: Enables tare.
Standard serial output if the weight value
overflows and is unstable.
0: Disables output.
1: Enables output.
.
Page 45
Default
value
2
n
0
1
20000
2
0. 0
0. 0
1. 0
2
1
1
1
User
setting
AD-4408A