Setting List; Basic Functions - A&D AD-4408A Instruction Manual

Weighing indicator for field networks
Table of Contents

Advertisement

11. Setting List

When performing maintenance, use the following list as a memo.
In case of inquiry about the product, inform of the user setting.

11.1. Basic Functions

Function No.
Parameter or
Setting range
fncf01
F
000000
to
111111
fncf02
0
6
to
fncf03
1
3
to
fncf04
0
3
to
fncf05
0
9
to
fncf06
1
10
to
fncf07
1
2
to
Description
Disable key switch
0: Does not disable the key switch function.
1: Disables the key switch function.
Assigns a function to the F key.
0: None
1: Manual print command
2: Hold
3: Alternate switch
4: Momentary switch
5: Clear the tare weight
6: Clear the zero value
Display update rate
1: 20 times/second
2: 10 times/second
3: 5 times/second
Assigns a function to the
0: None
1: Zero tracking in progress
2: Alarm (Zero range setting error, over)
3: F key status
Digital filter (cutoff frequency)
0: None
1: 11.0 Hz
2: 8.0 Hz
3: 5.6 Hz
4: 4.0 Hz
5: 2.8 Hz
6: 2.0 Hz
7: 1.4 Hz
8: 1.0 Hz
9: 0.7 Hz
Frequency divider ratio
Hold
1: Normal hold
2: Peak hold
display.
Page 43
Default
User
value
setting
000000
(binary)
0
1
0
8
1
1
AD-4408A

Advertisement

Table of Contents
loading

Table of Contents