Section 12-Electronics; Repositioning The Joystick - Invacare Pronto M91 Owner's Manual

With surestep
Hide thumbs Also See for Pronto M91:
Table of Contents

Advertisement

SECTION 12—ELECTRONICS
SECTION 12—ELECTRONICS
After any adjustments, repair or service and before use, make sure that all attaching
hardware is tightened securely - otherwise injury or damage may result.
Before performing any maintenance, adjustment or service verify that On/Off
switch on the joystick is in the Off position.

Repositioning the Joystick

NOTE: For this procedure, refer to FIGURE 12.1.
NOTE: Take note of position and orientation of mounting hardware for reinstalling the joystick 
assembly.
1. Turn the adjustment lock lever to release the joystick mounting tube from the 
mounting bracket.
2. Remove the joystick from the wheelchair.
3. Remove the three hex mounting screws, spacers and locknuts that secure the 
mounting bracket to the three mounting holes on the arm frame.
NOTE: The mounting bracket is mounted to the inside of the arm frame.
4. Reposition the mounting bracket on the opposite arm frame.
5. Using the three hex mounting screws, spacers and locknuts secure the mounting 
bracket to the three mounting holes of the arm frame.
6. If necessary, perform the following to reposition the adjustment lock:
A. Slide the adjustment lock from the mounting bracket.
B. Rotate adjustment lock 180° and slide adjustment lock over the opposite end of the 
mounting bracket.
7. Slide joystick mounting tube through the mounting bracket to the desired position 
and secure adjustment lock to tube by turning lever on adjustment lock.
NOTE: Joystick not shown.
Hex Mounting Screws
Adjustment Lock Lever
Pronto® M91™ with SureStep®
WARNING
Mounting
Bracket
FIGURE 12.1 Repositioning the Joystick
68
Locknuts
Arm Frame
Spacers
Joystick Mounting
Tube
Part No 1141450

Advertisement

Table of Contents
loading

Table of Contents