Calculating Formula For Limit Line; Bulk Cable And Connector Impedance - Agilent Technologies E5071C Compliance Tests

Esa displayport phy compliance test standard version 1 revision 2b agilent method of implementation (moi) for displayport cable compliance tests
Hide thumbs Also See for E5071C:
Table of Contents

Advertisement

Agilent MOI for DisplayPort Cable & Connector Compliance Tests
1. Press
Channel Next
should be set.
2. Press
Analysis
> Limit Test > Edit Limit Line to display the limit table shown below
(Initially, no segments are entered in the limit table). Using the limit table, create/edit a
segment.
3. Enter the limit line data following the tables below.
4. Click Return.
5. Click Limit Line and turn it ON.
6. Click Limit Test and turn it ON.
7. Repeat 1 to 6 for each Measurement items.

6.5. Calculating formula for Limit Line

6.5.1. Bulk Cable and Connector Impedance

6.5.1.1.
Impedance Profile Through Full-size DisplayPort Connector
Segment
Fixture
key and
Trace Next
Impedance Profile
Differential
Maximum
Impedance Value
Tolerance
100 ohm
+-10 %
key to activate the trace on which limit lines
Comment
Fixture should have trace lengths of no
more than 50 mm (2-inches)
39