Filter Screen - Teledyne 3110 Operating Instructions Manual

Portable trace oxygen analyzer
Table of Contents

Advertisement

Portable Trace Oxygen Analyzer
1. Use the UP/DOWN keys to navigate to the SPAN screen.
Then press ENTER to start the span calibration.
Once the span calibration has begun the screen changes to reveal a
span finish selection.
2. When the FINISH selection appears, press ENTER to end the
span calibration. The analyzer will accept the calibration and
return to the HOME screen.
Note: After successfully performing a span calibration, you are
returned directly to the HOME screen rather than back to
the previous menu.

3.2.15 Filter Screen

The 3110 includes user adjustable filter. The filter has settings 1-
10. Setting 1 is the least amount of filtering and 10 is the highest level of
damping. The filter is used to reduce the noise level of the O
More filtering is required for lower trace ranges such as 10 ppm. Adding
filtering will slow down analyzers response to changing O
lowest effective level of filtering should always selected. The filter level
should be lowered to 1or 2 when setting the span to avoid delays. For a
0-10 ppm range a filler setting or 6 or 7 should typically provide good
results.
Teledyne Analytical Instruments
Operation
readings.
2
levels. The
2
35