Power for the control module is provided by a center-tapped trans-
former which takes the 115 VAC input, reduces it to 40 VAC, and feeds the
voltage to a DC power supply identical to the one installed in the detector
module. Power supply outputs are +24 VDC unregulated, and +15 VDC
regulated.
Power
Supply
Power
Power
Transformer
115
VAC
60
Hz
Manual
Input
Fine
Zero
Current
Signal
I-to-E
Input
Converter
From
Analysis
Unit
115
VAC
60
Hz
Figure 2-5.
T o Timer
15 VDC
24 VDC
To
I-to-E
Converter
Span
Adjust.
Ext.
Volt.
Amplifier
Auto
Zero
Timer
Control Module - Block Diagram
Teledyne Analytical Instruments
A Business Unit of Teledyne Electronic Technologies
Operational Theory 2.0
Meter
Indication
Voltage
(0
to
1
V
f.s.)
Millivolts
(0
to
5
mV
Manual
Setpoint
Adjust.
#2
#1
Alarm
Relays
Comparator
E-to-I
(4
Converter
Output
f.s.)
N.O.
Com.
N.C.
N.O.
Com.
N.C.
Current
to
20
mA)
Standard
Valve
Control
Signal
2–11