Adjusting the Temperature Offset

If the meter is in a temperature measurement function,
use the following procedure to set an offset for your
temperature probe:
1.
Turn the rotary switch to temperature (
2.
Connect the temperature probe and probe adapter to
the
and V inputs on the meter.
3.
Place the temperature probe and an accurate
thermometer in a lag bath (i.e., a container with an
isothermal liquid).
4.
Press
to enter the Setup mode and
temperature adjust.
The primary display shows the measured value for the
temperature probe. This value is already adjusted by any
previously stored offset (shown in the secondary display.)
See Figure 5-1.
).