The ABPM6100 is intended for use in the electromagnetic environment specified below. The customer or the user
of the ABPM6100 should assure that it is used in such an environment.
Immunity test
±6 kV contact
Electrostatic discharge
(ESD)
±8 kV air
IEC 61000-4-2
Electrical fast
±2 kV for power supply
transient/burst
lines
IEC 61000-4-4
±1 kV for input/output
lines
±1 kV differential mode
Surge
IEC 61000-4-5
±2 kV common mode
Voltage dips, short
<5 % U
interruptions and
(>95 % dip in U
voltage variations on
for 0,5 cycle
power supply input
lines
40 % U
(60 % dip in U
IEC 61000-4-11
for 5 cycles
70 % U
(30 % dip in U
for 25 cycles
<5 % U
(>95 % dip in U
for 5 sec
Power frequency (50/60
3 A/m
Hz) magnetic field
IEC 61000-4-8
NOTE U
is the a.c. mains voltage prior to application of the test level.
T
DIR 80012720 Ver. B
ABPM 6100 Hardware Manual
Electromagnetic Immunity
IEC 60601
Compliance level
test level
±6 kV contact
±8 kV air
n.a.
n.a.
n.a.
T
)
T
T
)
T
T
)
T
T
)
T
3 A/m
Electromagnetic environment
guidance
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be
at least 30 %.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of the ABPM6100 requires
continued operation during
power mains interruptions, it is
recommended that the
ABPM6100 be powered from an
uninterruptible power supply or
a battery.
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.