Troubleshooting - Welch Allyn ABPM 6100 Hardware Manual

Hide thumbs Also See for ABPM 6100:
Table of Contents

Advertisement

* Always return ABPM 6100 monitor to an authorized service center for repairs.
Unauthorized service will void all warranties.
8.2

Troubleshooting

The following table contains a list of troublesome scenarios accompanied by suggestions
for problem solving.
Trouble
Cycle starts but cuff will not fully inflate
Blood Pressure readings are not displayed
during Regular Runs.
Patient activation button does not initiate
readings while in the Regular Run mode.
Time Regular Runs do not initiate
Blood Pressure readings are failing with
error codes being displayed
DIR 80012720 Ver. B
ABPM 6100 Hardware Manual
Response to Trouble
Ensure a connection with the ABPM 6100
monitor is secure and check bladder for
leak; replace if needed. Replace batteries
and try again.
Verify that the display setting is ON in the
Basic Parameters menu.
Verify that Start Button is ON in the Basic
Parameters menu.
Make sure that the time period is not set for
MAN (manual) operation.
Refer to the Error Codes section ABP
Workstation Software manual.