ABPM 6100 Hardware Manual
Welch Allyn
4341 State Street Road
Skaneateles Falls, NY
13153-0220 USA
www.welchallyn.com

Advertisement

Table of Contents
loading

Summary of Contents for Welch Allyn ABPM 6100

  • Page 1 ABPM 6100 Hardware Manual Welch Allyn Welch Allyn Staalweg 50 4341 State Street Road 2612 KK Delft Skaneateles Falls, NY The Netherlands 13153-0220 USA www.welchallyn.com 0297 DIR 80012720 Ver. B...
  • Page 2 ABPM 6100 Hardware Manual Copyright © Copyright 2008, Welch Allyn. All rights are reserved. No one is permitted to translate, reproduce or duplicate this manual or any part thereof, in any form, without prior written permission from Welch Allyn. Welch Allyn assumes no responsibility for any injury to...
  • Page 3: Assistance And Parts

    ABPM 6100 Hardware Manual Warranty, Service, and Spare Parts Warranty All repairs on products under warranty must be performed or approved by Welch Allyn. Unauthorized repairs will void the warranty. In addition, whether or not covered under warranty, any product repair shall exclusively be performed by Welch Allyn certified service personnel.
  • Page 4 ABPM 6100 Hardware Manual Limited Warranty Statement Welch Allyn, Inc. warrants that the Welch Allyn CardioPerfect Workstation computer based product you have purchased meets the labeled specifications of the Product and will be free from defects in materials and workmanship that occur within 1 year after the date of purchase.
  • Page 5: Table Of Contents

    ABPM 6100 Hardware Manual Table of contents OVERVIEW ...........................6 ABPM 6100 MONITOR ......................8 Introduction..........................8 Operation..........................9 INTERFACE ........................11 Preliminary Directive......................11 Prior to Hook-up ........................11 Sizing..........................11 Traditional Cuff Placement ....................12 Sleeve Cuff Placement......................12 Office Readings .........................14 Regular Runs........................14 Data ...........................15 KEYSPAN USB TO SERIAL ADAPTER INSTALL............17 MAINTENANCE AND CLEANING ..................20...
  • Page 6: Overview

    ABPM 6100 Hardware Manual 1. Overview The purpose of this manual is to provide an overview of information that you will need to safely and effectively use the ABPM 6100 monitor. Information about the ABP Workstation Software is available in the ABP Workstation software manual. Make sure that you familiarize yourself with all of the safety precautions listed in this chapter before attempting to use either device or the software.
  • Page 7 ABPM 6100 Hardware Manual IMPORTANT WARNINGS The following warnings apply to the ABPM 6100 Ambulatory Blood Pressure Monitors: DO NOT use the ABPM 6100 monitor in the presence of flammable Warning anaesthetics due to risk of explosion. DO NOT immerse the ABPM 6100 monitor in any fluid, place any fluid on the monitor, or attempt to clean either monitor with liquid detergents or cleaning agents.
  • Page 8: Abpm 6100 Monitor

    ABPM 6100 Hardware Manual 2. ABPM 6100 Monitor Introduction The ABPM 6100 Unit is worn by the patient in a belt or shoulder strap and is connected to a cuff, which is wrapped around the non-dominant arm. ABPM 6100 inflates the cuff at pre-...
  • Page 9: Operation

    ABPM 6100 Hardware Manual Operation This section provides a brief overview of the ABPM 6100 unit, how to load the batteries, and the unit controls. The ABPM 6100 features a simple design. One Start/Stop button functions as the primary control. The LCD (liquid crystal display) displays easy-to-read information. The ABP Workstation Software allows you to program the unit before the study and retrieve data after the study.
  • Page 10 ABPM 6100 Hardware Manual At this point the unit will be ready to upload a BP Study. When the unit is turned on subsequently, the unit will display the following: Three quick audible beeps. The number of BP readings (if any) in memory along with a flashing printer symbol for 3 seconds.
  • Page 11: Interface

    ABPM 6100 Hardware Manual 3. Interface This section describes how to connect the ABPM 6100 monitor to the patient for monitoring sessions. Preliminary Directive Advise the patient to: Wear a loose fitting blouse or shirt. Avoid wearing long sleeved sweaters or dresses during monitoring period.
  • Page 12: Traditional Cuff Placement

    ABPM 6100 Hardware Manual Arm Circumference Soft disposable Reusable sleeve Traditional reusable Cuff Size cuffs cuffs sleeve cuffs Part 5082-94-3 Part 6100-10 Pediatric Cuff 15.8-21.3 cm 16.0 – 21.8 cm Part 5082-95-3 Part 101340 Part 6100-11 Small Adult Cuff 20.0-27 cm 18-27 cm 21.1 –...
  • Page 13 ABPM 6100 Hardware Manual Sleeve cuff Step 1: To determine the correct cuff size for your patient, follow these simple steps: 1. To find the right sized cuff, wrap the cuff around the patient’s upper arm without sliding the arm through the sleeve.
  • Page 14: Office Readings

    ABPM 6100 Hardware Manual Do not remove the cuff between readings • Office Readings Once the ABPM 6100 monitor has been properly attached to the patient and the PC interface cable has been removed from the unit, it is important and necessary to perform office readings in order to make sure that the unit is functioning properly and does not cause discomfort to the patient.
  • Page 15: Data

    ABPM 6100 Hardware Manual PATIENT DO’S AND DON’TS: Finally, familiarize the patient with the following list of do’s and don’ts: • Wear a loose fitting blouse or shirt. • Record time, symptom/mood and activity/position in the patient diary. • Activate a reading (if the Start/Stop button is activated) at the first sign of symptoms.
  • Page 16 ABPM 6100 Hardware Manual See the ABP Workstation software manual for instructions on using the software and the data handling and reporting options it offers. DIR 80012720 Ver. B 16 / 30...
  • Page 17: Keyspan Usb To Serial Adapter Install

    ABPM 6100 Hardware Manual 4. Keyspan USB to Serial adapter Install If your computer doesn’t have a COM port you need to install the Keyspan Highspeed USB to serial adapter (USA 19H). The purpose of this section is to assist with the Install of the Keyspan Highspeed USB to Serial Adapter (USA19H) and determining the COM port assigned.
  • Page 18 ABPM 6100 Hardware Manual 4. Click “Next” 5. Click “Yes” if you agree with the license agreement. 6. Click “Next” 7. Click “Finish” 8. Insert the Keyspan Adapter into an available USB port. 9. The driver will load automatically after few seconds. When the light on the adapter flashes at a constant rate (~1 sec interval) the device is ready.
  • Page 19 ABPM 6100 Hardware Manual 11. Navigate as shown above and click “Keyspan Serial Assistant” 12. In the example above “COM4” was assigned to the adapter. Make note of which COM port yours assigned and set this as the COM port for connection to your device in the CardioPerfect Workstation.
  • Page 20: Maintenance And Cleaning

    ABPM 6100 Hardware Manual 5. Maintenance and Cleaning Preventative maintenance should be routinely performed to ensure the safe and efficient operation of the ABPM 6100 monitor. In addition, the monitor should be cleaned after each use. Maintenance The following inspections of the ABPM 6100 unit should be performed on a regular basis: •...
  • Page 21: Calibration Check

    ABPM 6100 Hardware Manual 6. Calibration Check Welch Allyn recommends that the calibration of the ABPM 6100 monitor be verified annually by the user using the following procedure: 1. With the ABPM 6100 unit already powered up, remove one of the AA batteries from the battery compartment.
  • Page 22 ABPM 6100 Hardware Manual • Welch Allyn recommends using the most sensitive pressure standard possible when performing calibration checks. A Setra Pressure Meter 2270-01, which is calibrated for ± 0.1 mm Hg, works well for this application.In the Federal Republic of Germany: In accordance with Med BetreibV, the ABPM must be returned every 2 years to instrumentation control for calibration certification.
  • Page 23: Specifications

    ABPM 6100 Hardware Manual 7. Specifications This section provides the unit specifications for the ABPM 6100 ambulatory blood pressure monitor. Power Requirements: • Two “AA” alkaline batteries or high capacity rechargeable batteries (NimH). Dimensions: • 124 x 70 x 33 mm (approximate).
  • Page 24 ABPM 6100 Hardware Manual Operator Control: • 1-button control and LCD. Automatic Measurement Intervals: • Programmable up to four separate time periods at 5 to 120 minute intervals. DIR 80012720 Ver. B 24 / 30...
  • Page 25: Troubleshooting And Error Codes

    ABPM 6100 Hardware Manual 8. Troubleshooting and Error Codes Error Codes The ABPM 6100 monitor displays error codes whenever an error situation is encountered. Error codes will display on the unit’s LCD. Error codes that apply to a specific reading will also display in the ABP Workstation Software when the data is read from the unit.
  • Page 26: Troubleshooting

    ABPM 6100 Hardware Manual * Always return ABPM 6100 monitor to an authorized service center for repairs. Unauthorized service will void all warranties. Troubleshooting The following table contains a list of troublesome scenarios accompanied by suggestions for problem solving. Trouble...
  • Page 27: Guidance And Manufacturer's Declarations

    ABPM 6100 Hardware Manual 9. Guidance and Manufacturer’s Declarations CAUTION The ABPM6100 needs special precautions regarding EMC and needs to be installed and put into service according to the following EMC information provided. Portable and mobile RF communications equipment can affect the ABPM6100.
  • Page 28 ABPM 6100 Hardware Manual Electromagnetic Immunity The ABPM6100 is intended for use in the electromagnetic environment specified below. The customer or the user of the ABPM6100 should assure that it is used in such an environment. Immunity test IEC 60601...
  • Page 29 ABPM 6100 Hardware Manual Electromagnetic Immunity The ABPM6100 is intended for use in the electromagnetic environment specified below. The customer or the user of the ABPM6100 should assure that it is used in such an environment. Immunity test IEC 60601 Compliance level –...
  • Page 30 ABPM 6100 Hardware Manual Recommended separation distances between portable and mobile RF communications equipment and the ABPM6100 The ABPM6100 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the ABPM6100 can help prevent electromagnetic interference by...