At this point the unit will be ready to upload a BP Study. When the unit is turned on
subsequently, the unit will display the following:
Three quick audible beeps.
-
The number of BP readings (if any) in memory along with a flashing printer symbol for 3
-
seconds.
One long audible beep.
-
Displays flashing time for 20 seconds (after 20 seconds the units turns itself off and
-
goes into Sleep Mode to conserve battery life)
The LCD Display
The LCD displays the following information depending upon the state that the unit is in:
-
The time is displayed whenever the unit is in Normal Mode and ready for an action.
-
A sun symbol is displayed whenever the buzzer is on (usually during the day).
A Crescent Moon symbol is displayed when the buzzer is off (usually during
-
programmed sleeping periods).
A clock is displayed whenever the unit is in Study Mode.
-
A battery symbol is displayed whenever the battery voltage is low and the batteries
-
need to be replaced.
Care needs to be taken to avoid removing the batteries from the unit for any
length greater than 15 minutes. Even though the battery level may be too low to
actuate the pumps, there will be enough charge to maintain the internal clock
and patient readings for many months.
When the ABPM 6100 is not being used for more than 3 months, care needs to
be taken to remove the batteries. Before the batteries are removed, download all
data or the data will be lost.
-
A printer symbol indicates that the unit contains readings in the memory.