Kodak I810 - Scanner Firmware - PC Installation Planning Manual page 7

I800 series scanners
Hide thumbs Also See for I810 - Scanner Firmware - PC:
Table of Contents

Advertisement

Line voltage and
frequency
Voltage requirements
AC power quality and
conditioning
4
Acceptable operational voltage:
100-127 V, 6.5 A, 50/60 Hz
200-240 V, 3.5 V, 50/60 Hz
The scanner is shipped with a package of line cords to accommodate all
worldwide installations.
CAUTION: Make certain nothing obstructs or is placed on power
cords. Do not use extension cords; they can cause
voltage loss. This may result in unreliable equipment
performance or equipment damage.
Four outlets are required:
One individual dedicated branch AC circuit employing two standard
3-prong, grounded duplex outlets. These outlets are designated for
the scanner and service diagnostic equipment.
The outlets should meet the following standards:
Amperage: 15 amp
Wire size: 12-gauge
Phase: 1 (single)
Receptacle: NEMA5-15R IG
NOTE: Receptacles should be placed within 5 feet (1.5 meters) of the
equipment.
Impedance between neutral and ground wires: less than 2 ohms
Neutral to ground voltage: less than 1 volt AC maximum
Voltage drop across circuit breaker: less than 0.1 volts
Impulse voltage spikes and transients: less than 50% of nominal line
voltage
Surge voltage: less than 5% of nominal line voltage
In some locations, even though recommended wiring practices have
been followed, the power reaching the scanner may exhibit unstable
power quality characteristics. System errors and intermittent data loss
could result.
Power conditioning equipment may be required to overcome the power
quality problems. Kodak can make arrangements for an on-site power
analysis in an area with exceptionally unstable power.
Consult with your electrical contractor or utility company to ensure that
on-site wiring complies with the specifications and that wall outlets have
isolated grounds and comply with codes.
A-61176 November 2001

Advertisement

Table of Contents
loading

Table of Contents