1.10 Error Indications

Scale errors

Only input encoder scale errors are reported by the ND 1200. Scale
errors are indicated by the presence of graphic bars across the DRO
screen instead of numbers. Scale errors can be caused by a few
conditions shown in this table:
Possible cause
Corrective action
Damaged encoder
Repair or replace the encoder.
reader head
Misaligned
Recalibrate the reader head.
encoder reader
head
Electrical noise on
Verify that the earth ground provided by the
the encoder input
ND 1200 power source is in good condition
and is connected to the power system
ground.
Verify that the encoder cable is shielded
and that the shield is connected to ground
at the ND 1200 end.
Verify that the reader head is not generating
electrical noise.
Stage (encoder)
Restrict stage velocity. If the velocity
velocity too high
required for error-free operation is slow,
recalibrate the reader head.
Faulty encoder
Repair or replace the encoder cable or seek
input connection
the assistance of your Metronics
distributor.