EXCELL PRECISION CO., LTD. Table of Contents INSTRUCTION FOR USE......................4 PREPARING TO USE THE BALANCE ..................4 PROTECTION IN THE TRANSPORTATION................5 CHAPTER ONE PRODUCT FEATURES..................6 1-1 Features ..........................6 1-2 Specifications ........................7 1-3 Plastic Shield Installation & Dimension ................9 1-4 Description of Power Supply ....................
Page 3
EXCELL PRECISION CO., LTD. 3-1-9 Zero Tracking Range Setting ............26 3-2 Weight Calibration and Gravity Compensation ..........27 3-2-1 External Weight Calibration ............. 28 3-2-2 Gravity Compensation ..............29 3-2-3 Restore to the Default Setting ............30 3-3 RS232 SERIAL INTERFACE SETTING ..........
Page 4
EXCELL PRECISION CO., LTD. 4-4-2 CLEAR EEPROM DATA ..............48 4-4-3 RESOULTION KEY SETTING ............49 4-4-4 GRAVITY CALIBRATION SETTING ..........50 4-5 LOCAL GRAVITY VALUE SETTING ............51 4-6 TEMPERATURE CALIBRATION ............53 4-6-1 SETTING THE TEMPERATURE CALIBRATION VALUE .......... 54 4-6-2 CHECK THE TEMPERATURE CALIBRATION VALUE ..........
EXCELL PRECISION CO., LTD. INSTRUCTION FOR USE Thank you for purchasing BH3 precision balance. To enable you to use this balance correctly, we suggest that you read these instruction carefully. 1. It must be prevented from getting wet. If it gets wet, please wipe it dry with a cloth. If there is something wrong with the machine, please contact your supplier.
EXCELL PRECISION CO., LTD. PROTECTION IN THE TRANSPORTATION Before using the BH3 Precision Balance, please read the following illustrations to take off the protection screw. For the maintenance of this machine, install the protection screw before the transportation to avoid any damage on the sensitive components.
EXCELL PRECISION CO., LTD. CHAPTER ONE PRODUCT FEATURES 1-1 Features 1/30,000 and 1/60,000 display resolution available. Auto calibration. Weighing, counting and percentage modes. Setting the usable units from 13 weighing units. High resolution switching function to display 10 times divisions.
EXCELL PRECISION CO., LTD. 1-4 Description of Power Supply Power supplies 1. 4 pcs of UM-2 dry batteries 2. 9V DC 300mA adaptor Power Consumption 1. The power consumption is around DC 0.5 mA (standby mode). 2. The power consumption is around DC 30~35 mA (system + load cell).
EXCELL PRECISION CO., LTD. 1-6 Display Stable Low battery Accumulation Zero Minus sign Tare Mode Unit display 1-7 Keyboard Function : Off key. Press the key to switch the balance off. : Function key. Press the key to select different operation modes: weighing mode, MODE counting mode, and percentage mode.
EXCELL PRECISION CO., LTD. CHAPTER TWO OPERATION INSTRUCTIONS 2-1 Zero Function →0← When the display shows weight value without anything placed on the platter, press the to re-zero the balance. The zero indicator is on, indicating a center-of zero gross weight condition.
EXCELL PRECISION CO., LTD. 2-3-1 Weighing Mode Ⅰ. Units Selection Press the key to select the desired weighing unit. UNITS There are 13 units available: g, ct, lb, oz, dr, GN, ozt, dwt, MM, tl.J, tl.T, tl.H and t. Ⅱ Precision Shifting Function...
EXCELL PRECISION CO., LTD. 2-3-2 Counting Mode Ⅰ. Sampling 1. Press the MODE key to select the Counting mode and the display shows pcs. 2. Press the key to select a sample size from 10, 20, 50, 100, or 200 pcs.
EXCELL PRECISION CO., LTD. 2-3-3 Percentage Mode I. Sampling 1. Press the key to select the “Percentage” mode and the display shows MODE 2. Press the key to select 100.0% or 100.00%. 3. Place enough sampling weight on the platter and press the key.
EXCELL PRECISION CO., LTD. CHAPTER THREE FUNCTION SETTINGS MODE MODE Press key for 3 seconds or hold keys together, the balance enters the function setting mode, and the display shows Please refer to page 16 ~ 25 General function setting...
EXCELL PRECISION CO., LTD. 3-1 GENERAL FUNCTION SETTING There are nine functions settings available Press ⇒ Back to the previous level ⇒ First unit setting ⇒ Operation units setting ⇒ Auto-off timer setting ⇒ Backlight mode setting ⇒ Noise filter setting ⇒...
EXCELL PRECISION CO., LTD. 3-1-1 First Unit Setting The default setting is “g” Use the key to select the unit in cycle, and press the key to confirm the setting. There are 13 weighing units available for selection: g, ct, lb, oz、dr, GN, ozt, dwt, MM, tl.J, tl.T, tl.H, t...
EXCELL PRECISION CO., LTD. 3-1-2 Operation Units Setting There are 13 weighing units available for selection. Use the or the key to switch between ON and OFF for each unit. The default setting for all 13 units are “ON” BH3 Series...
EXCELL PRECISION CO., LTD. 3-1-3 Auto-Off Timer Setting The default setting is “ 0 ”. Use the or the key to select one from the four modes: Press the key to confirm and the setting procedure is completed. ⇒ Auto-off not available ⇒...
EXCELL PRECISION CO., LTD. 3-1-4 Backlight Mode Setting The default setting is “ AUTO ” ( The auto backlight is on) Auto backlight ⇒ When there is weight on the platter (the value is larger than 10 divisions), or when any key is pressed, the backlight is on.
EXCELL PRECISION CO., LTD. 3-1-5 Noise Filter Setting Set a parameter in 01 to 02, according to the environment where the balance is positioned. ♦ Please select “ 01 ”, when operation environment is unstable. ♦ Please select “ 02 ”, when operation environment is stable.
EXCELL PRECISION CO., LTD. 3-1-6 Forward Zero Display Setting This function is not accessible for approval models. The default setting is “ 1 ” ( 1d, d=division ) Use the or the key to select a zero display range. Press the key to confirm the setting.
EXCELL PRECISION CO., LTD. 3-1-7 Backward Zero Display Setting This function is not accessible for approval models. The default setting is “ 2 ” ( 2d, d=division ) Use the or the key to select a proper parameter. Press the key to confirm the setting.
EXCELL PRECISION CO., LTD. 3-1-8 Auto Unit Weight Averaging Setting The default setting is “ Yes” (Automatic Unit Weight Calibration) Use the or the key to select “Yes” or “No.” Press the key to confirm the setting. ⇒ YES (...
EXCELL PRECISION CO., LTD. 3-1-9 Zero Tracking Range Setting Zero Tracking Range Setting Not available for approved model. ⇒ 3/4 d ) ( default value: Press key to select parameter. ⇒ 1/4 d ⇒ 2/4 d ⇒ 3/4 d ⇒ 1 d ⇒...
EXCELL PRECISION CO., LTD. 3-2 Weight Calibration and Gravity Compensation Press the ⇒ Weight Calibration ⇒ Gravity Compensation ⇒ Restore to the default setting This function is not accessible for approval models. When set as “YES” for the external Gravity calibration function, function is available.
EXCELL PRECISION CO., LTD. 3-2-1 External Weight Calibration The display shows the calibration weight Press the (equal to the full capacity) and the weight unit. The first digit flashes. The calibration weight can be Ensure the platter is on the modified if necessary.
EXCELL PRECISION CO., LTD. 3-2-2 Gravity Compensation Press the to show the history data 1 sec The display shows the latest modification done before 1 sec The display shows the 5 modification done before 1 sec The display shows the 4...
EXCELL PRECISION CO., LTD. 3-2-3 Restore to the Default Setting Press the Press the ⇒ Return to upper level ⇒ Restore to the default setting in external weight calibration and gravity compensation ⇒ Restore to the default setting in the general function...
EXCELL PRECISION CO., LTD. 3-3-1 Transmission Mode Setting The default setting: “ ” Use the or the key to select from the 4 modes available: Press the key to confirm the setting. ⇒ Directive Mode ⇒ Stable transmission + Directive Mode ⇒...
EXCELL PRECISION CO., LTD. 3-3-2 Baud Rate Setting The default setting: “ ” Use the or the key to select from the 7 baud rates: Press the key to confirm the setting. ⇒ 600 bps ⇒ 1200 bps ⇒ 2400 bps ⇒...
EXCELL PRECISION CO., LTD. 3-3-3 Communication Protocol Setting The default setting: “ ” Use the or the key to set the parameter Press the key to confirm the setting. ⇒ 8 data bit, No parity bit, 1 stop bit ⇒ 8 data bit, No parity bit, 2 stop bit ⇒...
EXCELL PRECISION CO., LTD. 3-3-4 Output Format Setting The default setting is “ ” Use the or the key to select format mode. Press the key to confirm the setting. ⇒ Format 1 ⇒ Format 2 ⇒ Format 3 ⇒ Format 4...
EXCELL PRECISION CO., LTD. 3-3-6 Weight Range Setting The default setting:“ ” (The weight range change will not be defined) Weight Range can be set 00 ~ 99. And “ 00 ” is default. When range is over set value, and Stable Transmission or Manual Transmission Modes is set the value will print out.
EXCELL PRECISION CO., LTD. 3-3-7 Date & Time Display Setting This function is available with RTC card. The balance will work with either dry batteries or the adaptor. When the balance is out of battery with its adaptor unplugged, time will need to be reset. Please replace the battery before proceeding to reset the time.
EXCELL PRECISION CO., LTD. 3-3-8 Date & Time Setting This function is available with RTC card. The balance will work with either dry batteries or the adaptor. When the balance is out of battery with its adaptor unplugged, time will need to be reset. Please replace the battery before proceeding to reset the time.
EXCELL PRECISION CO., LTD. CHAPTER FOUR SERVICE MODE ACCESS 1. To access the service mode, set the Mini Jumper SWA1 to the ADJ position (EEPROM UNLOCK), and then switch on the scale, and the display shows 2. The Mini Jumper SWA1 must be set back to the LOCK position when the calibration is completed.
EXCELL PRECISION CO., LTD. 4-1 CAPACITY CONFIGURATION Pres ↵ key Input “max. capacity + 9d“ Pres ↵ key A ⇒ Division Parameter: 1 ⇒ the last digit of weight value is “1” Parameter: 2 ⇒ the last digit of weight value is “2”...
EXCELL PRECISION CO., LTD. 4-2 WEIGHT CALIBRATION The display shows the maximum capacity Press ↵ key as the calibration weight which could be Press ↵ key changed if needed. The display shows The calibration weight should be no the maximum capacity as the less than 1/2 of full capacity.
EXCELL PRECISION CO., LTD. 4-3 LINEARITY CALIBRATION After the linearity calibration, the weight calibration should be done for one more time. The internal value of weight calibration should be within the range of the linearity calibration internal value. The maximum points of linearity calibration are 9 points which includes the zero point; at least 2 points should be set.
Page 45
EXCELL PRECISION CO., LTD. If the multiple value of weight is not Continue necessarily input, put the weight of 3rd point and the display shows the internal value. Input the multiple value of weight for the 3 point. The multiple value of weight...
Page 46
EXCELL PRECISION CO., LTD. Continue Press ↵ key to memorize the point The display shows the internal value after 1 second. →0← Press key to save all the value of linearity calibration. Or press key to abandon all the value of linearity calibration.
EXCELL PRECISION CO., LTD. 4-4-1 APPROVAL PARAMETER SETTING Press ↵ key Press Press Press Press Press Press Press Press Press Press ↵ key key to select the parameter and press ↵ key to finish the approval parameter setting. Press MODE key to exit the setting and go back the main menu and the display shows ⇒...
EXCELL PRECISION CO., LTD. 4-4-2 CLEAR EEPROM DATA Press ↵ key Press Press Press Press Press Press Press Press ↵ key key to select the parameter and press ↵ key to finish the approval parameter setting. Press MODE key to exit the setting and go back the main menu and the display shows ⇒...
EXCELL PRECISION CO., LTD. 4-4-3 RESOULTION KEY ( ) SETTING Press ↵ key Press Press Press Press Press Press Press ↵ key key to select the parameter and press ↵ key to finish the approval parameter setting. Press MODE key to exit the setting and go back the main menu and the display shows ⇒...
EXCELL PRECISION CO., LTD. 4-4-4 GRAVITY CALIBRATION SETTING Press ↵ key Press Press Press ↵ key key to select the parameter and press ↵ key to finish the approval parameter setting. Press MODE key to exit the setting and go back the main menu and the display shows ⇒...
EXCELL PRECISION CO., LTD. 4-5 LOCAL GRAVITY VALUE SETTING Manufacture gravity value is in the 00 batch. Local gravity could be input for 9 times. To input the local gravity value in function, the new input value would be saved as the 1 batch data and clear all the history data.
Page 53
EXCELL PRECISION CO., LTD. Press ↵ key to show the data history After 1 sec The display shows the last time number and after 1 sec. shows Press Press the local G value. After 1 sec The display shows the last 2 time number and after 1 sec. shows...
EXCELL PRECISION CO., LTD. 4-6 TEMPERATURE CALIBRATION Press ↵ key Press Press Press Press Press Press Press Press ⇒ Go Back to Upper Level (Exit Temperature Calibration Mode) ⇒ Setting the Temperature Calibration Value ⇒ Check the Temperature Calibration Value ⇒...
EXCELL PRECISION CO., LTD. 4-6-1 SETTING THE TEMPERATURE CALIBRATION VALUE Press ↵ key to check the current temperature Ensure there is no load on the platter and press ↵ key to wait for stable to check the current temperature The display shows the full capacity, put the weight on platter and press ↵...
EXCELL PRECISION CO., LTD. 4-6-2 CHECK THE TEMPERATURE CALIBRATION VALUE Press ↵ key key to check the span of 0℃. Press Press key to check the span of 25℃. Press Press key to check the span of 40℃. Press ↵ key key to check the temperature value of every calibration point and use key to check the span value of that point.
EXCELL PRECISION CO., LTD. 4-6-3 DELETE THE TEMPERATURE CALIBRATION VALUE Press ↵ key Press ↵ key to check the span of 0℃. Delete Press Press key to Press ↵ key check the span of 25℃. Delete Press Press Press ↵ key key to check the span of 40℃.
EXCELL PRECISION CO., LTD. APPENDIX I UNIT CONVERSION TABLE [ MET.CARAT ] 0.2 g [ AVOIRDUPOIS POUND ] 453.59237 g oz [ AVOIRDUPOIS OUNCE ] 28.349523125 g [ AVOIRDUPOIS DRAM ] 1.7718451 g GN [ GRAIN ](U.K) 0.06479891 g ozt [ TROY OUNCE ] 31.1034768 g...
EXCELL PRECISION CO., LTD. APPENDIX III RS232 BI-DIRECTION FORMAT General Format Gross + SP 1 6 SP SP SP g 6 SP SP c Tare 6 SP SP k CR LF Overload O L + SP SP SP SP SP SP SP SP SP SP SP SP...
Page 62
EXCELL PRECISION CO., LTD. RS232 OUTPUT FORMAT Format 1 (RS1 04 is set form 1 ) HEAD1 HEAD2 SIGN UNITS HEAD1 (3 byte) Stable Unstable Overload HEAD2 (3byte) Gross Weight Net Weight SIGN (1byte) Positive weight Negative weight DATA (8byte)
Page 63
EXCELL PRECISION CO., LTD. Format 2 (RS1 04 is set form 2 ) CR LF SIGN DATA UNITS SIGN (1byte) Positive weight Negative weight DATA (8byte) 8 BYTES including the decimal point UNITS(4byte) CR,LF (2byte) Total length of transmission is 15 bytes ( 1 + 8 + 4 + 2 = 15 )
Page 64
EXCELL PRECISION CO., LTD. Format 3 (RS1 04 is set form 3 with option card RTC) Weight Mode CR LF CR LF CR LF CR LF CR LF CR LF CR LF CR LF Counting Mode CR LF CR LF...