4-4-1 APPROVAL PARAMETER SETTING

Press ↵ key
Press
Press
Press
Press
key
Press ↵ key
Use
or
and press ↵ key to finish the approval parameter setting.
Press MODE key to exit the setting and go back the main menu and the display shows
BH3 Series
key
key
key
key to select the parameter
.
⇒ No regulation (The zero point at switch on has no limits.)
⇒ OIML regulation (The internal value of zero point at switch on should be at the
range of calibration zero point ±10% full scale. The internal value of Zero key and
zero tracking should be at the range of zero point at switch on ±2% full scale.)
⇒ Sri Lanka regulation 1 (The internal value of zero point at switch on should be at
the range of calibration zero point ±3% full scale. The internal value of Zero key
and zero tracking should be at the range of calibration zero point ±3% full scale.)
⇒ Sri Lanka regulation 2(The internal value of zero point at switch on should be at
the range of calibration zero point ±3% full scale. The internal value of Zero key
and zero tracking should be at the range of zero point at switch on ±3% full scale.)
Press
Press
Press
key
Press
key
Press
key
,