10.9
Electromagnetic compatibility
Noise immunity/interference emissions
Electrostatic discharge on casing parts (ESD)
RF irradiation
Burst on power cables
Burst on signal lines
Surge on power cables
Magnetic fields
•
Constant fields
•
Alternating fields
Voltage fluctuations
Line reaction to harmonics
Line reaction to voltage fluctuations
Limits of radio disturbance characteristics
See also
Guidance and manufacturer's declaration – electromagnetic emissions
RadiForce LX600W
Instruction for Use, 11/2013
10.9 Electromagnetic compatibility
•
EN 60601‑1‑2 Class B
•
EN 55011:2007 + A2:2007
•
FCC/ICES: FCC part15 subpartB/ICES003
•
C-Tick:
AS/NZS CISPR22:2006,
AS/NZS 61000‑3‑2:2007 + A1:2009
•
VCCI/JEIDA: CISPR22:2005,
JIS C61000‑3‑2:2005
EN 61000-4-2
8 kV air, 6 kV contact
EN 61000-4-3
80 MHz ... 2 500 MHz,
3 V/m 80 % AM 1 kHz
EN 61000-4-4
2 kV
EN 61000-4-4
1 kV
EN 61000-4-5
1 kV symmetric, 2 kV unsymmetrical
EN 61000-4-8
4000 A/m (maximum)
1000 Aeff/m (maximum)
EN 61000-4-11
EN 61000-3-2:2006 + A1:2009 + A2:2009
GB17625.1
EN 61000-3-3
GB9254
Technical specifications
[}
42]