Connecting; Safety Information For Connection - Eizo RadiForce LX600W Instructions For Use Manual

8mp 60.1" lcd monitor
Table of Contents

Advertisement

5

Connecting

5.1

Safety information for connection

All safety information and warnings for the device must be observed to ensure danger-free
operation.
CAUTION
Changes to device
Do not make any mechanical or electric changes to the device.
EIZO GmbH will not be held liable if changes are made to the device.
CAUTION
Shielding measures
Follow all shielding measures according to the country-specific EMC Directive. If these
guidelines are not observed, device malfunction may result.
CAUTION
Grounding measures
To ensure that the permissible leakage current is not exceeded in the case of a fault, you
must ground the device with a separate ground line.
CAUTION
Excessive currents, short circuits, and ground faults
In accordance with national standards and regulations, protection against excessive cur‐
rents, short circuits, and ground faults must be incorporated into the building installation.
NOTICE
Changes to device settings
Device settings may only be adjusted by trained service personnel. Otherwise, the warran‐
ty is void.
NOTICE
Disconnecting from line power
Always set the power switch to "Off" before disconnecting the device from power. Other‐
wise the device could be destroyed.
RadiForce LX600W
Instruction for Use, 11/2013
5.1 Safety information for connection