Excell SUBe AirCharge Service Manual page 27

Waterproof weighing scale
Table of Contents

Advertisement

EXCELL PRECISION CO., LTD.
NOTE 6
(o) (p) ( q ) (r ) (s ) (t)
(o)  Minimum division setting
Parameter description:
Decimal system:
Input 1, 2, or 5 as the
minimum division for the
weight value
(p)  Decimal point position
Parameter description:
Decimal system:
0 
1 
0.0
2 
0.00
3 
0.000
4 
0.0000
5  0.00000
(q)  Weight UNITS displayed
Parameter
UNITS
Notation
system
symbol
 If multi-range or multi-interval is set, the weight UNITs oz, GN, viss, ct shall not be used.
(r)  Scale change point (Input the parameter 0, 1, 2, or 3)
Parameter description:
0  full segment range
1  full segment range
2  2 segment range (changes at 1/2 of full scale)
3  3 segment range (changes at 1/6 of full scale and 2/3 of full scale)
(s)  Save the weight UNITS at preferred slots (no more than the number of set
weight UNITS)
Parameter description:
1  the first slot (the weight calibration UNITS)
2  the second slot
3  the third slot
4  the fourth slot
5  the fifth slot
(t)  0: multi-interval
020359XX
0
0
1
0
1
0
1
2
4
kg
g
lb
lb,oz
1
10
10
10
16
kg
g
lb
lb
1
1
2
2
2
2
1: multi-range
4
Hexadecimal notation system:
Input 1, 2, 4, or 8 as the
minimum division for the
weight value
Hexadecimal notation system:
0 
1 
2 
3 
5
6
oz
GN
10
10
Icon 7
Icon 6
indication
Indication
25
0.0
0.0.0
0.0.00
0.0.000
7
8
dwt
ct
10
10
Icon 6
Icon 6
indication
indication
ZSME300000434

Advertisement

Table of Contents
loading

This manual is also suitable for:

Sube